SALD-2300

Availablity: In stock

Catalog/Part Number: SALD-2300
Brand: SHIMADZU
Supplier: SHIMADZU
Supplier Location: JAPAN
Inventory Weight: unavailable
Unit of Measurement:
Inventory Brochure: 1571820932.pdf |

Catalog/Part Number:: SALD-2300

Sorry no inventory found for this brand at this time.

Description

Using the SALD-MS23 sampler, the measurement range is 17nm to 2500μm for wet measurement.
For example, PSL particles with a median diameter of 50nm and stainless balls with a diameter of 2mm can be measured by a single analyzer.

By enabling measurement under wide particle concentration conditions (0.1ppm to 20%), changes in particle size distribution depending on particle concentration can be evaluated.
Previously, the particle concentration of a sample had to be adjusted to meet the optimum conditions of analyzers by dilution or concentration using a centrifuge. In these cases, changes in particle size distribution, such as agglomerations or dispersions, could not be considered.

Dispersions and agglomeration can be caused by dilutions.
In some cases, dilutions can accelerate dispersions, but in other cases, they can create agglomerates. To ensure optimum, the initial particle concentration must be determined without dilutions or concentrations. After the evaluation of particle size distribution at the initial state, the effects of particle concentration from dispersions and agglomerations must be evaluated.

Wide particle concentration range must be covered to evaluate the dissolution process of sample particles.
This is necessary because the progress of dissolution makes the particle concentration low in comparison to the first particle concentration.

SALD-2300 can measure particle size distribution under the conditions of particle concentration from 0.1ppm to 20%.
When the sampler SALD-MS23 or batch cell SALD-BC23 is used, measurements are possible under conditions of concentration from 0.1ppm to 100ppm.
When the high-concentration sample measurement system SALD-HC23 is used, high-concentration samples up to 20% can be measured because the negative effects of multiple scattering are prevented.

Catalog/Part Number: SALD-2300 Category:

Technical Specifications

Measurement principle

Laser Diffraction Method

Measurement range

SALD-MS23: 17nm(0.017μm) to 2500μm

SALD-BC23: 17nm(0.017μm) to 400μm

SALD-HC23: 30nm(0.030μm) to 280μm

SALD-DS5: 300nm(0.3μm) to 2500μm

Note 1: The measurement range varies according to the shape etc. of the particle.

Measurement Unit: SALD-2300 (P/N: 347-61700-42[115V], 347-61700-44[230V])

Light source

Red Semiconductor laser (Wavelength 680 nm)

Light detector

Detector elements for UV semiconductor laser
Total 84 elements (78 forward, 1 side, 5 back)

System Compliance

Class 1 Laser Product, CE compliant

Required power supply

115 or 230 VAC as ordered 100 VA

Dimensions & weight

W680mm×D280mm×H430mm, 31kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Note 2: Reference sample and USB cable (2 m) supplied as standard

Sampler: SALD-MS23 (P/N: 347-61701-42[115V], 347-61701-44[230V])

Dispersing bath

Capacity: 100280cm3

Sonicator

Frequency about 32 kHz, output about 40 W

Liquid Pump

Radial pump, maximum flow rate 2000cm3/min

Liquid Pump Material

Stainless (SUS 304, SUS 316), Tetrafluoroethylene (PTFE), Perfluoroelastmor (FEP) or Kalrez, Thermoflon Pascal (inside)

Liquid Supply Pump

Diaphragm pump, maximum flow rate 750cm3/min

Liquid Supply Pump Material

Tetrafluoroethylene, polyvinyl dene fluoride

Flow Cell

Quartz glass

Required power supply

115 or 230 VAC as ordered, 200 VA

Dimensions & weight

W390mm×D520mm×H430mm,18kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Note 3: USB cable (2 m) supplied as standard

Batch Cell: SALD-BC23 (P/N: 347-61702-42)

Cell Material

Quartz glass

Required Liquid Volume

Approx. 12 cm3

Stirrer Mechanism

Up-and-down movement of blade

Dimensions & weight

W100mm×D120mm×H140mm, 0.8kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

High-concentration Sample Measurement System: SALD-HC23 (P/N: 347-61703-42)

Cell Material

Borosilicate glass

Required Liquid Volume

Approx. 0.15 cm3

Dimensions & weight

W20mm×D100mm×H9mm, 0.2kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Cyclone Injection Type Dry Measurement Unit: SALD-DS5
(P/N: 347-61706-42[115V], 347-61706-44[230V])

Sample suction types

Cyclone type / One shot type / Hand shot type

Suction nozzle

Can be selected from 3 types

Sampling Unit Specifications

System

Cyclone type

Communication method

USB (PC control)

Required power supply

115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor)

Dimensions & weight

W 240×D 310×H 210 mm, 10 kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Pressure Regulator Specifications

Primary pressure

0.6 to 0.8 MPa

Secondary pressure

0.05 to 0.5 MPa

Filtration rating

Removing particles of 5 μm or larger

Connection to air source

Tube of 6 mm outside diameter

Communication method

USB (PC control)

Required power supply

115/230VAC(±10%), 100VA, 50/60Hz (excluding dust collector and compressor)

Dimensions & weight

W 130×D 223×H 233 mm, 3 kg

Operating Environment

Temperature: 10 to 30°C, Humidity: 20 to 80 % (no condensation)

Requirements for compressor and dust collector

Compressor

Output

0.4 kW

 

Minimum pressure

Approx. 7 kgf/cm2 (approx. 0.69 MPa)

Air discharge

45 L/min

Tank volume

Approx. 30 L

Dust collector

Type

Vacuum cleaner (Paper pack type)

Dust collecting efficiency

More than 99% for 0.3 μm particles

Capacity

2.0 m3/min or more

Vacuum

Approx. 2000 mm Aq or lower

Suction hose diameter

Approx. 32 mm

Software Specifications

WingSALD II

Measurement and Data Display Functions

Measurement of Particle Size Distribution

Allows measurements using measurement assistant function (interactive process based on SOP)

Refractive Index Setting

Automatic refractive index calculation function (LDR method: Light Intensity Distribution Reproduction Method) makes setting the refractive index easy.

Real-Time Display

Particle size distribution/light intensity distribution simultaneous display

Diagnostics/Adjustments

Self-diagnostic function and cell check function

Recalculation of Particle Size Distribution

Batch recalculation of max. 200 distributions

Display of Particle Size Distribution Data

Displays overlay of max. 200 distributions

Display of Light Intensity Distribution

Displays overlay of max. 200 distributions

Statistical Data Processing

Max. 200 sets of data (also allows overlaying max. 200 data sets)

Time-Series Processing

Max. 200 sets of data

3-Dimensional Graphing

Max. 200 sets of data

Data Transfer via Clipboard

[Image Output]: Outputs entire data sheet or graph only.
[Text Output]: Outputs summary data, particle size distribution data, or light intensity distribution data.

Data Sorting

Sorts by file name, sample ID, sample number, or refractive index

Output Conditions

Particle Size (μm) Divisions

Fixed 51 or 101 divisions    : User settable 51 divisions

Particle Amount (%) Divisions

Fixed 51 divisions : User settable 51 divisions

Distribution Basis

Count, length, area, or volume

Expression of Cumulative Distribution

Oversized or undersized

Expression of Frequency Distribution

q, q / Δ×, q / Δlog ×

Smoothing Levels

10 levels

Distribution Function Fitting

Rosin-Rammler distribution, logarithmic Gaussian distribution

Data Shifting

±10 levels

Report Function

Single data sets (6 templates), overlaid data (5 templates), statistical data, time-series data, or 3D data can be selected and output using batch processing

Data Analysis Functions

Scattering Angle Evaluation Function

Evaluates scattering characteristics within micro angle regions for samples such as optical films and sheets.

Data Emulation Functions

Emulates measurement results from other instruments and measurement principles, using SALD series measurement results.

Mixture Data Simulation Function

Simulates particle size distributions using any mixture ratio of multiple particle size distributions.

Data Connection Function

Combines two particle size distributions with different measurement ranges at any particle size point to create a single particle size distribution.

Continuous Measurement Function

Continuously measures changes in particle size distributions and particle diameters over time, at intervals as short as one second, and saves the results.

Note 4:The LDR method automatically calculates an appropriate refractive index based on consistency between the actual measured light intensity distribution and one reproduced (recalculated) from particle size distribution data. This method was developed by Shimadzu and published in two technical papers. It is sometimes called the "Kinoshita Method", in academic communities, after the name of Shimadzu’s engineer.

PC Requirements

The software is included standard on a CD-R with the SALD system (optical system).
Install the software on a PC that meets the following specifications.

OS

Windows 7