Description
A specific volume (100-mL) of test sample is introduced between the two concentric electrodes in the Cell. Using a triaxial cable, the Cell is connected to a sensitive dc ammeter in the Console and a dc voltage is applied between the two electrodes. Due to the physical geometry of the Cell, the Cell constant is stable and calculable. Thus, the resultant peak current is readily transformed and digitally displayed as picosiemens per meter (pS/m), also commonly known as conductivity units (CU). As defined in ASTM Standard Test Method D 4308, a CU is equal to a picosiemen per meter (1 CU = 1 pS/m). The enclosure technique employed in the Cell precludes the need for external screening from stray electrical fields.
TEST METHOD:
For the determination of the “rest” electrical conductivity of aviation fuels and other similar low-conductivity hydrocarbon liquids in the range from 0.1 to 2000 pS/m.
FEATURES:
Capable of
being modified to provide higher conductivity ranges
Optional
capability to measure sample temperature
Mechanical
design of the Conductivity (KSLA) Cell allows accurate self-measuring of the
test sample volume (100 mL)
Electronic
circuitry automatically samples and holds the reading
Large five
digit liquid crystal display
Internal self-check
calibration of all electronic circuits
Ease of
disassembly and cleaning of the Conductivity Cell
Completely self-contained powered by 6 standard 9 volt batteries monitored by a low battery indicator.
INCLUDED ACCESSORIES:
Electronics
Assembly
Batteries
Console
Instruction
manual on CD
KSLA Fuel Cell
Certificate of
Calibration
Interconnecting Cable Assembly.
Technical Specifications
Conforms to
the specifications of:
ASTM D4308
0 – 2000 pS/m
0 – 20,000
pS/m
0 – 200,000
pS/m
Zero(0) Reviews