Automatic Cleveland Open Cup Flash Point Tester
Catalog/Part Number:: K87400
Description
Test Method
For flash and fire point of all
petroleum products, except fuel oils and those having an open cup flash below
79°C (175°F).
Features
Conforms to ASTM D92 and related
specifications
Simple automation routine for
easy operation
Flash point operation between
ambient and 400°C
Flash and Fire Point detection by
Ionization Ring
Gas or electric ignition
Data Storage: 200 results
Automatic gas cut-off at the end
of the test
Equipped with Emergency Safety
Switch
Description
The automated Cleveland Open Cup flash
point tester accurately determines flash and fire point temperatures of viscous
petroleum products including oils and bitumens over an extended temperature
range. When examining highly viscous specimens, a preheating time and
temperature are set in order to liquefy the sample for testing. The surface
skin from bituminous samples can be removed with a skimmer. The flash/fire
point tests are simply conducted by mounting the flash cup filled with sample
into the test position and selecting a pre-programmed test method or the search
mode to determine an approximate flash point. The test results are
automatically corrected to standard pressure (101.3 kPa). Equipped with a
differential Pt-100 RTD probe, the system is designed to duplicate the response
time of a mercury-in-glass thermometer. Multiple sensors continually monitor
instrument function, displaying an error message if a problem is detected. The
performance of the ionization sensor which detects the flash and fire points is
continuously monitored, and the user is notified upon the need of replacement.
If a flash is not detected 20°C above the expected flash point or at 420°C,
then the test is automatically aborted for safety. The system is easily
interfaced with an external PC for operation and method updates. When
performing a test, the system will display the stirring speed, temperature
curve (also printed out), and current test status. The system alerts the user
if the first application of the ignitor results in a flash or if no flash point
is detected at the end of the test program. If a flash is not detected 30°C
above the expected flash point or at 400°C, then the test is automatically
aborted for safety.
Included Accessories
Ticket Printer
Test Cup
PT100 Probe
Detection Cable
Gas Tubing
RS232C Output
Data Acquisition Software
Technical Specifications
Specifications
Conforms to the specifications
of: ASTM D92; IP 36, 403; ISO 2592; AFNOR T60-118; EN 22592; DIN 51376
Electrical Requirements
115V 60Hz 1000W
230V 50Hz 1000W
Dimensions
lxwxh,in.(cm):
21x10.5x19.75(53.5x26x50)
Net Weight: 53 lbs (24 kg)
Zero(0) Reviews